Alarms - continued
Message in display What it means / Alarm How to make it stop
*
No key pressed in xx:xx
hours. Pump will stop and
alarm in 30 seconds
Q
This is an
insulin pump
Pump is stopped
Medical help may
be required
!
Battery
low
Change battery
soon
!
Bolus
Interrupted
Remaining
Amount: x.xxu
Duration: xx:xxhr
Continue?
Q
Delivery will stop in 30 seconds
unless you press Cancel. Attention
alarm.
This is the Automatic Off alarm.
Continuous alarm.
Press Cancel. This is the notification of the
Automatic Off alarm. If you do not press Cancel,
the pump will activate the Automatic Off alarm
in 30 seconds.
If you are able, press OK. This indicates that no
keys or buttons were pressed for an extended
period (you select the interval when setting the
feature). If you don’t want to use the Automatic
Off feature, you can turn it off (see page 28).
*Using the Treatment Assistant
PC Communications System, you can create your own custom message for this alarm.
If the screen has been customized, the message will be different than the one shown above.
Alerts
Message in display What it means / Alarm How to make it stop
This alert occurs only during waking
hours (6:00AM to 10:00PM). A
series of 4 Attention alarms given
every 4 hours until battery is dead.
Your last bolus was interrupted. The
amount not delivered (and remain-
ing duration, if applicable) appear in
the display. Attention alarm.
Press OK. Change the battery as soon as pos-
sible.
To deliver the rest of the bolus, press Yes. If you
do not want to deliver the rest of the bolus, press
No.