Using the Pump - Basics
43
6. Set site change reminder and restart delivery
The Site Change reminder screens appear here if turned
on in Pump Settings / Personalize / Alert (see page
28). Setting the site change reminder will cause the
pump to beep (or vibrate) to remind you that it is time
to change your infusion set and access site.
☞
Warning: You should not change your access site
right before bed, as you would be unaware of any
problems with the site. You must test your BG level
1 to 2 hours after starting a new cartridge to make
sure insulin is being properly delivered.
Fill
Cannula
< Back Fill
u
0.0
Cozmo
<Stop Menu>
Basal: Pattern 1
Rate: 1.150 u/hr
Correction Bolus
Temporary Rate
Basal Pattern
Load
<Home Select>
Cartridge
Menu
<Done Select>
Load
Fill tubing
Fill cannula
Reminder
;
;
;
;
Press Select
Press
Done
Press
Home
At reset, choose Ye s *. Choose interval
between alerts, then press Next;
choose time, then press Next
Site Reminder
Reset: Yes
Interval: 3 days
Time: 6:00PM
<Back Next>
Cartridge
Menu
<Done Select>
Load
Fill tubing
Fill cannula
Reminder
;
;
;
*When you choose Yes, the alert clock
is reset to zero, and will begin a new
countdown (in this example, in 3 days
at 6:00 PM your pump will give the
alert). Choosing No allows the alert
clock to continue the countdown from
the last time it was reset.