Using the Pump - Basics
38
FastForward
Please wait...
FastForward
Please wait...
Not ready
Pushrod must move
forward.
Remove cartridge
then press OK
OK
No cartridge
detected
Cartridge missing or
overfilled
Install cartridge to
confirm sensor
Cancel Confirm
OK
Sensor
Confirmed
Now pushrod must
move forward
Remove carridge
then press OK
*In order for the pump to assure that the cartridge is correctly connected
to the pushrod, the pushrod must start in a forward position. If it is
not far enough forward when you press Load, or if you attempt to load a
cartridge that contains less than 20 units of insulin, the screens shown
at left will appear. Remove the cartridge, then press OK. The pushrod moves
forward, then you return to the cartridge menu. If your cartridge contains less
than 20 units of insulin, you will need to start with a new filled cartridge. With
Load still chosen, press Select.
**If, during the loading process, you intentionally or accidentally remove the
cartridge, or if the cartridge is too full (more than 300 units), the screens
shown at left will appear. The pump needs to check to make sure that the
cartridge sensor is working properly.
• If the cartridge is not installed, install it;
• if a cartridge is installed make sure it is not too full (fill with no more than 300
units). Press down gently on the cartridge (this will push excess insulin into the
infusion set tubing). Press Confirm. The pump will confirm that the sensor is
working properly. (If the sensor is not working, you will get this message
repeatedly and your pump will need to be serviced). If the sensor is confirmed, you
must remove the cartridge and press OK. The pushrod moves forward, then you
return to the cartridge menu. With Load still chosen, press Select.