Deltec Cozmo Manual do Utilizador Página 118

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 117
Technical Information
106
ment 1 (1996; Amendment 2 (1996)
CISPR14 (1993) - International Special Committee on Radio
Interference, limits and methods of measurement of radio in-
terference characteristics of household electrical appliances,
portable tools and similar electrical apparatus. Amendment 1
(1996)
IEC 529 - Degrees of protection provided by enclosures (IP
Code)
RTCA/DO-160D (7/97) - Environmental conditions and test
procedures for airborne equipment: SECT 21 — Emissions of
radio frequency energy (Radiated Emissions Only — Category
M Limit)
UL2601-1 (1998) - Medical Electrical Equipment, Part 1: Gen-
eral requirements for safety
Specifications
General specifications (nominal)
Minimum increment of resolution:
0.005 Units per activation
Size:
3.2” × 1.8” × 0.95” (80mm × 47mm × 24mm)
Weight:
Approximately 3.2 ounces (90g) including battery and
cartridge
Classification:
Type CF
J
(protection from electric shock)
Moisture protection:
IPX8 (watertight to a depth of 12 feet [3.64 meters] for 3
minutes, or 8 feet [2.4 meters] for 30 minutes
Pump alarms/alerts:
Pump stopped, blockage, low battery, dead battery, low
cartridge volume, empty cartridge, system fault, delivery
limit
Optional alarms/alerts: end of temporary rate, tempo-
rary rate active, site change, check blood glucose, missed
meal bolus, automatic off
Maximum infusion pressure:
23 psi
Maximum time to blockage (occlusion) alarm:
Minimum Maximum
Bolus 5 u/hr 24 sec. 48 sec.
2 u/hr 1 hr. 3.9 hrs.
0.05 u/hr 37 hrs. 74 hrs.